Download Print this page

Changing The Battery - TESTO 175-T1 Instruction Manuals

175 series
Hide thumbs Also See for 175-T1:

Advertisement

3
4
5
info@Testo-Direct.com
1 Please read out the saved data before changing the battery. See
testo ComSoft
If it is not possible to read out the saved data due to low battery
capacity, please carry out the following:
2 Remove the screw at the back of the data logger using a small
1
crosstip screwdriver.
3 Using a screwdriver, lift the back wall at the bottom of the data
logger and then remove from data logger
4 Attach the jumper (included with spare battery) to the plug
connector beside the connection for the battery
The inserted jumper prevents the memory from being deleted or
written over.
2
5 Take the battery out of the battery compartment and pull out the
plug-in connection to the data logger
6 Connect the new battery to the data logger`s plug-in connection
and place in battery compartment.
Only original Testo spare batteries should be used
(see 12. A A ccessories/Spare p p arts, P. 34 for Part nos).
7 Remove the jumper from the plug connector.
8 Hold the back wall at a 45° angle to the top of the instrument
and then flip down.
Ensure that both O rings are positioned on the screw so as to
guarantee that it is sealed completely.
9 Press the back wall onto the logger with your thumbs. Make
sure that it is closed properly
10 Place the data logger in the desk-top holder and attach to
interface.
11 Start the
data logger by clicking twice on the required connection with the
left mouse button.
- The New battery window is opened
- The data logger is now ready for use.
www.
9 9 . . C C h h a a n n g g i i n n g g t t h h e e b b a a t t t t e e r r y y
software Instruction manual
First change the battery and then read out the saved data
(no data will be lost).
software and set up a connection to the
testo ComSoft
Enter the date of battery change.
Enter the temperature range in which you will be using the
data logger and confirm with OK.
Testo-Direct
1
.
.
2
.
3
4
and then secure using screw.
5
.
1.888.475.5235
.com
25

Advertisement

loading

This manual is also suitable for:

175-t3175-s1175-s2175-h1175-h2175-t2