Fluke 281 User Manual page 116

40 ms/s arbitrary waveform generators
Hide thumbs Also See for 281:
Table of Contents

Advertisement

281, 282, 284
Users Manual
The full procedure is as follows:
CAL 03
CAL 04
CAL 05
CAL 06
CAL 07
CAL 08
CAL 09
CAL 10
CAL 11
CAL 12
CAL 13
CAL 14
CAL 15
CAL 16
CAL 17
CAL 18
CAL 19
CAL 20
CAL 21
CAL 22
CAL 23
CAL 24
CAL 25
CAL 26
CAL 27
CAL 28
CAL 29
CAL 30
CAL 31
CAL 32
CAL 33
CAL 34
CAL 35
CAL 36
CAL 37
CAL 38
CAL 39
CAL 40
CAL 41
CAL 42
CAL 43
CAL 44
CAL 45
CAL 46
CAL 47
CAL 48
CAL 49
CAL 50
CAL 51
15-4
CH1 DC offset zero
CH1 DC offset at + full scale
CH1 DC offset at - full scale
CH1 Multiplier zero
CH1 Multiplier offset
CH1 Waveform offset
CH1 Output level at full-scale
CH1 20 dB attenuator
CH1 40 dB attenuator
CH1 10 dB attenuator
CH1 Not used
CH1 Not used
CH1 Not used
CH2 DC offset zero
CH2 DC offset at + full-scale.
CH2 DC offset at - full-scale
CH2 Multiplier zero
CH2 Multiplier offset
CH2 Waveform offset
CH2 Output level at full-scale
CH2 20 dB attenuator
CH2 40 dB attenuator
CH2 10 dB attenuator
CH2 Sum offset
CH2 SCM level at full-scale
CH2 AM level at full-scale
CH3 DC offset zero
CH3 DC offset at + full scale
CH3 DC offset at - full scale
CH3 Multiplier zero
CH3 Multiplier offset
CH3 Waveform offset
CH3 Output level at full-scale
CH3 20dB attenuator
CH3 40dB attenuator
CH3 10dB attenuator
CH3 Sum offset
CH3 SCM level at full-scale
CH3 AM level at full-scale
CH4 DC offset zero
CH4 DC offset at + full scale
CH4 DC offset at - full scale
CH4 Multiplier zero
CH4 Multiplier offset
CH4 Waveform offset
CH4 Output level at full-scale
CH4 20dB attenuator
CH4 40dB attenuator
CH4 10dB attenuator
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for -10 V ±3%
Adjust for minimum voltage (AC)
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ± 1 mV
Adjust for 0·1 V ±1 mV
Adjust for 2·236 V AC ±10 mV
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for -10 V ±3%
Adjust for minimum voltage (AC)
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ± 1 mV
Adjust for 0·1 V ±1 mV
Adjust for 2·236 V AC ±10 mV
Adjust for 0 V ±5 mV
Adjust for 5 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for -10 V ±3%
Adjust for minimum voltage (AC)
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ± 1 mV
Adjust for 0·1 V ±1 mV
Adjust for 2·236 V AC ±10 mV
Adjust for 0 V ±5 mV
Adjust for 5 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for -10 V ±3%
Adjust for minimum voltage (AC)
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ± 1 mV
Adjust for 0·1 V ±1 mV
Adjust for 2·236 V AC ±10 mV

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

282284