Information marked with this symbol indicates that the To use the device for its intended use, only use accessories and operating manual must be observed. spare parts which have been approved by Trotec. Operating manual – layer thickness measuring device BB30...
Improper use Information about the device Do not use the device in potentially explosive atmospheres, for measurements in liquids or at live parts. Trotec accepts no Device description liability for damages resulting from improper use. In such a The device is used to determine the coating thickness on case, any warranty claims will be voided.
Page 4
No. Designation Display Cross control with the buttons for up, down, right and left Detachable measuring head Sensor Battery compartment with cover Extension cable, device connection Extension cable, measuring head connection Power button Operating manual – layer thickness measuring device BB30...
0 to 40 °C (32 to 104 °F) at 20 to 90 % RH Power supply 2 batteries 1.5 V AAA Dimensions (length x 114 x 27 x 54 mm width x height) Weight 152 g Operating manual – layer thickness measuring device BB30...
1. Press the Power button (8) for approx. 2 seconds. ð The display will be switched on and the device ready for operation. Operating manual – layer thickness measuring device BB30...
Page 7
ð The measured value will be indicated on the display. ð If the calibration was successful, the result should be 0.0. ð OK and CANCEL will be indicated at the bottom of the display. Operating manual – layer thickness measuring device BB30...
Page 8
1. Press the Left button (2) to call up the main menu. 2. Use the Up or Down button (2) to select the menu item Measure Mode. 3. Confirm your selection by pressing the Left button (2). Operating manual – layer thickness measuring device BB30...
Page 9
(e.g. PC). 1. Press the Left button (2) to call up the main menu. ð The device will be displayed as BB30. 2. Use the Up or Down button (2) to select the menu item Set. 3. Confirm your selection by pressing the Left button (2).
Group (1) ... Group (50) Display of the measured values stored in the respective group. Calibration CAL Zero of FE CAL Zero of NFE DEL Zero of FE DEL Zero of NFE Operating manual – layer thickness measuring device BB30...
ð The measured values from the selected group are Device Address box. If Hex is activated, a sequence of stored in the Excel table. letters will be displayed. If Decimal is activated, a sequence of numbers will be displayed. Operating manual – layer thickness measuring device BB30...
Page 12
Data filtering selection Selecting a group for displaying measurement data Connect button Establishing the connection to the selected COM port COM port selection Selecting a COM port File menu Saving and loading data sets Operating manual – layer thickness measuring device BB30...
Repair Do not modify the device or install any spare parts. For repairs or device testing, contact the manufacturer. Operating manual – layer thickness measuring device BB30...
Page 14
Trotec GmbH & Co. KG Grebbener Str. 7 D-52525 Heinsberg +49 2452 962-400 +49 2452 962-200 info@trotec.com www.trotec.com...