Calibration 
 
 
Automatic Mode Calibration 
Set the meter to SLM mode, A weighting, High range, and Slow. 
Attach a 94.0dB (1000Hz) sound calibrator to the microphone. 
Enter SET mode and go to the parameter 94ADJ. 
 
Press the Run  
to appear. 
 
Press the Enter 
The meter should now read 94.0dB   (+/‐ 0.1dB). 
 
If the measured value does not equal 94.0 (+/‐ 0.1dB), run the 94ADJ routine again. 
Press the Enter 
 
nd Quality Products Online at:
 key to start the automatic offset routine and wait for the flashing dB number 
 key to save the entry and to get back to the measure window. 
 key to save the entry and to get back to the measure window. 