Recommendations To Other Types Of Insert - Presys T-350P Technical Manual

Dry block calibrators
Table of Contents

Advertisement

PRESYS
 Instruments

3.2 - Recommendations to Other types of Insert

When the user uses a different type of insert, other than the one used to adjust
the dry block, for example the cup like insert fulfilled with tiny steel balls, the accuracy
from the manual is no longer valid, since the thermal contact is different in relation to that
insert.
However, another control sensor adjustment can be performed to the new insert.
Do that using an external probe with high accuracy, placed in the new insert and follow
the steps described in item 7.2 - Probe Calibration.
This new readjustment of the calibrator indication overwrites the factory
calibration and it remains even after turning off the dry block.
Should you want the factory calibration back again, activate the option RESTORE
from menu calibration / probe.
T-350 P / T-650 P
Page 15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T-650p

Table of Contents