11.TECHNICAL DATA

Dimensions
• Depth
• Width
• Height
Weight
Temperature range
Height of figures
Power supply
Power consumption
• with wireless module disabled
• with wireless module activated
Maximum runtime in battery mode
• with wireless module disabled
• with wireless module activated
If you wish to check that the scales are properly calibrat-
ed, proceed as follows:
1. Switch the scales off if necessary.
2. Hold down any key and start up the scales.
The current content of the calibration counter will
flash in the display for a few seconds.
3. Compare the output content of the calibration
counter with the figure on the calibration counter
sticker.
The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, recali-
bration is necessary. Contact your service partner or the
seca customer service department. Should recalibration
be necessary, the seca calibration counter sticker
shown above will be replaced by the adjacent recali-
bration sticker to identify the calibration counter read-
ing. This sticker will be secured with an additional seal
by a person authorised to perform recalibration. The
recalibration sticker is available from the seca customer
service department under the number 14-05-01-886.
Technical data for
seca 376
358 mm
620 mm
190 mm
3.5 kg
+10° C to +40°C
21 mm
Battery
Power supply unit (optional)
approx. 32 mA
approx. 50 mA
approx. 5,600 minutes
Power supply unit recom-
mended
Technical data • 59
English

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents