rate and width are automatically varied in a
cycle pattern. The pulse width is decreased
by 50% from its original setting in 0.5 second,
then the pulse rate is decreased by 50%
from its original setting in 0.5 second.Total
cycle time is 1 second. In this mode, pulse
rate(2-150Hz) and pulse width(50-300 µs)
are fully adjustable.
The SD1(Strength-Duration) mode consists
of automatic modulation intensity and pulse
width in 40% range. The intensity is always
increasing while the pulse width is decreasing
and vice-versa. The intensity is decreased by
40% while the pulse width is increased by 40%
in 5 seconds. In the next 5 seconds, the intensity
is increased by 40% while the pulse width is
decreased by 40%.Total cycle time is 10
seconds. Pulse rate ( 2~150Hz) and pulse width
(50~300µs) are fully adjustable.
The SD2(Strength-Duration) mode consists
of automatic modulation intensity and pulse
width in 70% range. The intensity is always
increasing while the pulse width is decreasing
and vice-versa. The intensity is decreased by
70% while the pulse width is increased by 70%
in 5 seconds. In the next 5 seconds, the intensity
is increased by 70% while the pulse width is
decreased by 70%.Total cycle time is 10
seconds. Pulse rate( 2~150Hz) and pulse width
(50~300µs) are fully adjustable.
Constant stimulation based on setting value.
Only pulse width, pulse rate and timer are
adjustable in this mode. "Constant" is equal
10
http://www.pdffactory.com
to the "Normal" mode of a TENS unit.
20 Synchronous
Stimulation of both channels occurs
synchronously. The "ON" time including
Mode (S)
"Ramp Up" and "Ramp Down" time.
Therefore, the setting of ON Time should be
no less than two times of the "Ramp" time in
this mode.
ON TIME ≥ Ramp up + Ramp down
21 Alternate
The stimulation of the CH2 will occur after
Mode(A)
the 1st contraction of CH1 is completed. In
this mode, the setting of ON Time should be
no less than two times of the "Ramp" time.
The OFF Time should be equal or more than
the ON Time.
ON TIME ≥Ramp up + Ramp down
OFF TIME ≥ON TIME
22 Timer
Two Adjustable timers, from 1 to 60 minutes
or Continuous. Adjustable in 1 minute each
step from 1 to 15 minutes, and 5 minutes
each step from 15 to 60 minutes. Treatment
time countdown automatically.
23 Patient
This unit can store 60 sets of operation
Compliance Meter records. Total recorded time is 999 hours.
24 Low Battery
A low battery indicator will show up when
Indicator
the battery is low.
25 Operating
Temperature:0°~40°C
Condition
Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1060Hpa
26 Remark
There may be up to a +/-5% tolerance of all
parameters and +/-20% tolerance of
amplitude & voltage.