Cation Example Applications; Dionex Ionpac Cs12A Blank And Standard Analysis - Thermo Scientific Dionex ERS 500 Product Manual

Anion / cation electrolytically regenerated suppressor / carbonate eluents, 2 / 4 mm
Hide thumbs Also See for Dionex ERS 500:
Table of Contents

Advertisement

4
– Example Applications

4.7 Cation Example Applications

4.7.1

Dionex IonPac CS12A Blank and Standard Analysis

Sample:
Concentrator Column:
Guard Column:
Analytical Column:
Eluent:
Eluent Flow Rate:
Electrolytic
Suppressor:
or Chemical
Suppressor:
Chemical Regenerant:
Expected Background Conductivity:
Expected System Operating Back Pressure:
Figure 9
Figure 14A
Thermo Scientific
031956-11
100 µL of Standard or Blank neutralized in the Dionex CERS 500
Low Pressure Trace Cation Concentrator (i.e. Dionex TCC-XLP1, Item # 063889)
Dionex IonPac CG12A Guard Column
Dionex IonPac CS12A Analytical Column
20 mM Methanesulfonic acid
1.0 mL/min
Dionex Cation Electrolytically Regenerated Suppressor, Dionex CERS 500
Recycled Eluent Mode
Dionex Cation Chemically Regenerated Suppressor, Dionex CCRS 500
100 mN TBAOH
20 mM MSA: <1 µS
Without Guard: 1,400 psi (9.65 MPa)
With Guard: 1,850 psi (12.75 MPa)
Dionex IonPac CS12A Blank and Standard Analysis
Product Manual for the ERS 500 Suppressor
For Research Use Only. Not for Use in Diagnostic Procedures.
Analyte
Conc..
(µg/L)
1. Lithium
5.0
2. Sodium
20.0
3. Ammonium
25.0
4. Dimethylamine
B
5. Potassium
50.0
6. Trimethylamine
B
7. Magnesium
25.0
8. Calcium
50.0
Figure 14B
Page 55 of 84

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dionex aers 500Dionex cers 500eDionex aers 500 carbonateDionex cers 500Dionex aers 500e082540 ... Show all

Table of Contents