Cleaning Procedure - ADWA AD1000 User Manual

Ph/mv/ise/temperature meter
Table of Contents

Advertisement

For refillable electrodes:
Refill the electrode reference chamber with fresh
AD7082 electrolyte solution. Allow the electrode to
stand upright for 1 hour. Follow the storage procedure
above.

CLEANING PROCEDURE

Soak in deionized water for approximately 30 minutes.
IMPORTANT: After performing any of the cleaning
procedures, rinse the electrode thoroughly with deion-
ized water, refill the reference chamber with fresh elec-
trolyte and soak the electrode in AD70300 storage
solution for at least 1 hour before taking measurements.
71