GETTING STARTED Thank you for your purchase of this OWL certification test kit. Included in this kit is a WaveTester optical power meter and a WaveSource multimode fiber optic light source. This quick reference guide will assist you in setting up your test kit for a standards-based certification test.
Detector (connector port on light source may vary) WaveTester power meter Dual OWL multimode light source VERIFY PROPER OPERATION OF THE EQUIPMENT STEP 1 Connect the power meter and light source together with the first patch cable as shown in Figure 3.
STEP 5 Compare the reading on the WaveTester to Table 1 for acceptable power levels. The patch cable should be replaced if it is below the acceptable power level. Core Size Acceptable Power Level Replace at 62.5/125 µM -19.0 to -22.0 dBm -22.00 dBm 50/125 µM -22.0 to -25.0 dBm...
STEP 18 Connect the WaveTester and Dual OWL to the link under test as shown in Figure 5 below. 1 2 3 4 5 6 7 8 9 10 11 12 1 2 3 4 5 6 7 8 9 10 11 12...