OPERATING INSTRUCTIONS (CONT)
Using the Reweigh Function
The Reweigh function is used to reweigh the baby while he or she is still on the tray.
To reweigh, the baby should remain on the tray and a weight measurement should be on the display.
Simply press REWEIGH/ RECALL.
Using the Recall Function
The Recall function is used to recall the last weight measurement after the scale has shut down or
the display was cleared.
The scale automatically stores the last weight measurement. This value is stored until a new
measurement is taken or it is intentionally cleared by pressing ON/ZERO.
To recall a weight value after the baby's weight has been measured, the baby has been removed
from the tray and the display is off, press ON/ZERO and then REWEIGH/ RECALL. The stored
value will now display.
To clear the recalled weight and unlock the scale for more weighing, either press RECALL or
ON/ZERO.
Programming Auto-Off Time
The 2210KL / 2210KL4 has been pre-set at the factory to shut down after one minute of inactivity.
This can be changed to any increment (from 12 seconds to 25 minutes) to suit your needs.
1. Press ON/ZERO and wait for the display to show "0.0". Do not place anything on tray.
2. Hold the MENU button down and press the RECALL button. "Set-0" will be displayed along with
the up and down arrows; the "0" will be flashing.
button repeatedly until "Set-3" appears, press ENTER.
3. Press the
4. The Auto-Off time currently programmed in the scale will display as minutes and tenths of minutes
(i.e. "03.2"); the far left digit will be flashing. Use the
press ENTER to advance to the next digit. For example, setting the time to "00.5" will turn the scale
off after 30 seconds of no use; "02.0" will turn the scale off after 2 minutes; "25.0" will turn the scale
off after 25 minutes.
5. Press ENTER when complete to save and return to normal weighing operation.
and
ENTER
MENU
Figure 7: Menu button
11
buttons to set the desired value,