Honeywell SPM Flex User Manual page 25

Single point monitor gas detector
Hide thumbs Also See for SPM Flex:
Table of Contents

Advertisement

Tubing (optional)
Sample and exhaust tubing calculations
This table shows the flow rate, tubing length, transport time, and maximum
pressure and vacuum at the inlet and exhaust points�
Tubing lengths vary among gases� If the pressure or vacuum on the inlet/exhaust
lines does not meet the recommended values, the detector may encounter flow
faults�
Description
Tubing length, m (ft)
Transport time (sec)
Flow rate (cc/min.)
Inlet
Tubing OD in mm (in.)
Tubing ID in mm (in.)
Tubing Length, m (ft)
Outlet
Tubing OD, mm (in)
Tubing ID, mm (in)
The overall maximum load on the pump between the inlet and the exhaust should not exceed 10 inches H
NOTE
Use Teflon Fluorinated Ethylene Polymer (FEP) tubing to ensure proper sample
transport�
Optional external filters can be installed to reduce noise or to monitor for gas at
the location of the detector� To do this an inline filter is simply connected to the
gas inlet port� The area around the detector is then being monitored as opposed
to a sample being drawn from a remote location�
filter
John
Guest
fittings
Figure 16. External filter
Sample Specifications
Maximum
30 (100)
20 (66)
19
13
700-1200 (flow is set and controlled per calibration)
6.35 (0.25)
3.18 (0.125)
30 (100)
6.35 (0.25)
4.76 (0.188)
SPM Flex Gas Detector
10 (33)
0
7
1
O.
2
25

Advertisement

Table of Contents
loading

Table of Contents