Power Consumption; Ambient Temperature And Humidity; Heat Dissipation; Acoustic Noise Level - Beckman Coulter Navios Instructions For Use Manual

Flow cytometer
Table of Contents

Advertisement

Country
Europe and
other
applicable
countries
Japan

Power Consumption

<1500 Watts

Ambient Temperature and Humidity

Temperature fluctuations within the ambient temperature range can affect performance. For
optimal performance, Beckman Coulter recommends the ambient room temperature is maintained
between 15.5°C and 32°C (60°F to 90°F) and does not fluctuate more than 2.8°C (5°F) from the
ambient temperature at alignment. Keep the humidity between 30% and 85%, without
condensation.

Heat Dissipation

Heat dissipation is 720W (2457 Btu/hour). Provide sufficient air conditioning (refer to Ambient
Temperature and Humidity).

Acoustic Noise Level

 60db

Drainage

The waste line from the Cytometer is connected to a 20-L waste container. Dispose of the waste in
accordance with your local regulations and acceptable laboratory procedures.
773232AH
Dedicated Lines with Isolated Grounds
One dedicated line at 220 Vac, 50/60 Hz
at 10 A for the Supply Cart.
or
One dedicated line at 240 Vac, 50/60 Hz
at 10 A for the Supply Cart
One dedicated line at 100 Vac, 50/60 Hz
at 15 A for the Supply Cart.
Special Requirements
Non-Dedicated Lines
Two non-dedicated lines at 220 Vac,
50/60 Hz at 10 A - one for the tower
computer and a second, for the
monitor. A third non-dedicated line
is required if the optional printer is
being installed.
Three non-dedicated lines at
100 Vac, 50/60 Hz at 15 A - one for
the tower computer and a second,
for the monitor. A third non-
dedicated line is required if the
optional printer is being installed.
Installation
2
2-3