Parameter Button
Selecting this displays the Setup Parameter Screen.
Limits
The Limits grid allows you to specifyAcquisition Limits for Acquisition Time, Elapsed Time,
Maximum Events and Live Gate.
Set Acquisition/Elapsed Time
The 'Set Acquisition/Elapsed Time' dialog will allows you to set the required Acquisition and
Elapsed times for the current protocol.
The Acquisition Time must be less than or equal to the Elapsed Time. If no Elapsed Time is specified
then the Acquisition Time will be used. Both Acquisition and Elapsed Times must be between 10 and
99,999 seconds.
Acquisition Time
This field allows you to set Acquisition Time for the current Protocol.
Elapsed Time
This field will allows you to set the Elapsed Stop Time, which accounts for time in Pause mode, for
the current Protocol.
OK
Pressing OK will validate the Acquisition Time and Elapsed Time values. If one or more values is
invalid then you will be notified and the dialog will remain open. If both values are valid then the
dialog will close and the specified values will be indicated in the Cytometer Control dialog.
773232AH
Using Navios Software
10
Cytometer Menu
10-161
Need help?
Do you have a question about the Navios and is the answer not in the manual?