Levelflex FMP51, FMP52, FMP54 PROFIBUS PA
Endress+Hauser
Navigation path
Diagnostics → Event logbook → Event list
A maximum of 100 event messages can be displayed in chronological order.
The event history includes entries for:
• Diagnostic events
• Information events
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
–
: Event has occurred
–
: Event has ended
• Information event
: Event has occurred
Calling up and closing the remedial measures
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message about the remedial measures closes.
12.6.2
Filtering the event logbook
Using the Filter options parameter, you can define which category of event messages is
displayed in the Event list submenu angezeigt werden.
Navigation path
Diagnostics → Event logbook → Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information
12.6.3
Overview of information events
Info number
I1000
--------(Device ok)
I1089
Power on
I1090
Configuration reset
I1091
Configuration changed
I1092
Trend data deleted
I1110
Write protection switch changed
I1137
Electronic changed
I1151
History reset
I1154
Reset terminal voltage min/max
I1155
Reset electronic temperature
I1156
Memory error trend
I1157
Memory error event list
Diagnostics and troubleshooting
Info name
103
Need help?
Do you have a question about the Levelflex FMP51 and is the answer not in the manual?