Fluke 321 Calibration Information Manual page 9

32x series
Hide thumbs Also See for 321:
Table of Contents

Advertisement

Functional
Test
5520A Output
AC Amps
0.6 A @ 60 Hz
0.07 A @ 50 Hz
0.7 A @ 50 Hz
0.07 A @ 400 Hz
0.7 A @ 400 Hz
1 A @ 50 Hz
7 A @ 50 Hz
7 A @ 400 Hz
1 A @ 400 Hz
Ohms
30 Ohm
50 Ohm
350 Ohm
AC Volts
35 V @ 50 Hz
350 V @ 60 Hz
600 V @ 60 Hz
35 V @ 400 Hz
350 V @ 400 Hz
600 V @ 400 Hz
DC Volts
-350 V
35 V
350 V
600 V
Table 3. Performance Tests 322
Nominal
30 A @ 60 Hz
28.9
3.5 A @ 50 Hz
2.9
35 A @ 50 Hz
33.4
3.5 A @ 400 Hz
2.9
35 A @ 400 Hz
33.4
50 A @ 50 Hz
48.6
350 A @ 50 Hz
343.2
350 A @ 400 Hz
339.0
50 A @ 400 Hz
48.0
30 Ohms
29.2
Beeper must be on
50 Ohms
49.0
Beeper must be off
350 Ohms
346.0
35 V @ 50 Hz
34.0
350 V @ 60 Hz
345.3
600 V @ 60 Hz
592.3
35 V @ 400 Hz
34.0
350 V @ 400 Hz
345.3
600 V @ 400 Hz
592.3
-350V
-355.8
35V
33.9
350V
344.2
600V
590.5
Low Limit
31.1
4.1
36.6
4.1
36.6
51.4
356.8
361.0
52.0
30.8
51.0
354.0
36.0
354.7
607.7
36.0
354.7
607.7
-344.2
36.1
355.8
609.5
ClampMeters
Performance Tests
High Limit
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

322

Table of Contents