Validation

Some industries require sampling equipment to be validated before being
used. For the MicroBio MB1 this can be achieved using the MicroBio
Validation Kit, part number P0059 available from Cantium Scientific Limited
or your local distributor.

Calibration

It is recommended the MicroBio MB1 is regularly calibrated in accordance
with specific industry best practice.  Typically, this will be on the anniversary
of the instrument entering service.
The only calibration adjustment on the instrument is the flow rate, by default
set to a volumetric flow rate of 100 litres per minute. Calibration is normally
performed at an air pressure of 1013 mbar at 20ºC.
The normalised
conditions the unit is calibrated at are printed on the calibration label on the
rear end of the MicroBio MB1 and on the calibration certificate.
19