Cleaning

The sampling head should be sterilised preferably by autoclave, but in the
field disinfect with suitable sterilising wipes, such as IPA wipes or Tuffie 5
wipes (VernaCare part number 901SW225TY). Wipes may also be used for
general cleaning of the sampler body.
observe a disinfection regime between each sample. Always sample from
areas of lower contamination towards those of higher contamination. Wash
exposed skin and clothing before sampling and avoid drinking, eating and
smoking in the test area. All equipment should be handled aseptically.
The MicroBio MB1 is suitable for use with Hydrogen Peroxide Vapour (H
vapour) bio-decontamination. As the MicroBio MB1 has a fan within the
sampling head, it is recommended for optimum decontamination the
MicroBio MB1 is set to sample at maximum volume, allowing H
be drawn through the head, fan and exhaust.
Tests have been carried out by Bioquell (UK) Ltd on a range of sensitive
electronic equipment to determine effects of such processes. The conditions
of these tests were:
Throughout the tests, Bioquell demonstrated that bio-decontamination with
H
O
vapour does not appear to be detrimental in any way, affect operational
2
2
aspects or aesthetics of sensitive electronic equipment.
information on Bioquell H
Bioquell document BDS-3-MATCOMP-V3.2 (www.bioquell.com).
Gas Concentration
Conditioning time
Gassing time
Aeration time
Vapour sterilisation above the test conditions
levels and times will invalidate the warranty.
O
vapour material compatibility, please refer to
2
2
In areas of low bioaerosol levels,
1000 ppm
20 minutes
30 minutes
150 minutes
18
O
2
2
O
vapour to
2
2
For further