Diagnostics R-Net Lcd Control Panel - Permobil M5 Corpus User Manual

Hide thumbs Also See for M5 Corpus:
Table of Contents

Advertisement

User's Manual Permobil M5 Corpus
Event
The wheelchair switches itself off after a
certain period of inactivity (20-30 min).
The wheelchair stops while being driven.
The wheelchair can only be driven at
reduced speed.
The wheelchair cannot be charged.

Diagnostics R-net LCD control panel

General
When an error or a fault occurs in the wheelchair's electronics,
information about it is displayed in the control panel display. This
information can then be used to diagnose where the error, or fault,
occurred and its cause.
Possible cause
The electronics' energy saving mode has
been activated.
The cable connection to the control
panel has come loose.
Main circuit breaker tripped.
Seat lift or seat angle raised too high.
Applies only to an powered seat lift and
seat angle.
Main circuit breaker switched to off
position after e.g. battery replacement.
The charging circuit breaker has tripped.
Troubleshooting
Remedy
Switch the wheelchair on again using the
start button on the control panel.
Insert the cable in the control panel.
See page 243.
Lower the seat lift or seat angle.
See page 243.
Wait five minutes, the circuit breaker will
automatically reset.
261