To look at control and status signals during execution of a routine

To look at control and status signals during execution
of a routine
The state analyzer will trigger on the start of the routine whose control and
status signals are to be examined with finer resolution than once per bus
cycle. When it triggers, it will switch its "arm" level true. The timing analyzer
will trigger when it receives the true arm level and detects the transition
represented by glitch/edge1.
Select the state analyzer Trigger menu and define term R_START to
1
represent the starting address of the routine.
Under State Sequence Levels, enter the following sequence
2
specification:
While storing "anystate" TRIGGER on "R_START" 1 time
Store "anystate"
Select the timing analyzer Trigger menu.
3
Define the Edge1 term to represent a transition on one of the control
4
signals.
Under Timing Sequence Levels, enter the following sequence
5
specification:
TRIGGER on "arm • Edge1" 1 time