Battery Test Probes - Fluke BT520 User Manual

Battery analyzer
Hide thumbs Also See for BT520:
Table of Contents

Advertisement

BT520
Users Manual

Battery Test Probes

To connect test probes to the battery pole:
1.
Use the inner tip of the test probe to touch the target surface.
2.
Push the test lead to set-back the inner tip, until both the inner tip and the outer tip
are fully connected to the target surface. This will ensure a proper 4-wire
connection to the battery terminal.
Stable and correct readings are shown only when both the inner tip
and the outer tip of the test probe are fully connected to the battery
posts. To get more accurate battery internal resistance reading, do
not connect the test probes to screws. See Figure 3-2.
Examine for open fuse before mΩ measurement by connecting the
outer tips of both probes. If the mΩ reading changes from OL to
dashes and then backs to OL, the fuse is good. If mΩ reading
remains as OL, the fuse is open and needs a replacement.
In this function, the voltage between the positive and negative poles
of a battery must be < 60 V. A voltage >60 V causes the fuse to
open.
3-8
Note
Small Hole
Circle
Cylinder Clamp
Figure 3-2. Connect Test Probe to Battery
Cylinder Screw
Cube
Strip
hsz008.eps

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bt510

Table of Contents