Set Up The Sensors-Setup Wizard - Hach FL1500 User Manual

Flow logger
Hide thumbs Also See for FL1500:
Table of Contents

Advertisement

Push the UP and DOWN arrows to change the value. Push the LEFT and RIGHT arrows to move
the cursor.
Option
About
Status Display
Setup (slideshow)
Date and Time
Time zone
Export/Import
Display
Security
Language
Unit Preferences
Clear Data
Restore Factory
Defaults
Set up the sensors—Setup Wizard
Pre-requisites: Install the sensor in the process and the sensor cable in the logger before this task is
started.
The Setup Wizard is the easiest procedure to configure and calibrate the sensors. The Setup Wizard
menu prompts the user for information about the sensor and the flow channel, then calibrates the
sensor. As an alternative, the user can go into each menu item separately and enter the setup
information.
1. Select Sensor Setup>Change port assignments.
2. Select the number of the sensor connector in the instrument where the sensor wires are installed.
3. Select the sensor name. Select OK.
The sensor name shows next to the selected port number.
4. Select Setup Port [1] (sensor name).
5. Select Setup Wizard.
6. Select the options on each screen.
28 English
Description
Shows the instrument description, serial number and firmware version.
Sets the display to show the measurement screens in a slideshow format. Refer to
Show the data as a slideshow
Sets the time and date. Selects the format for the date and time. Options:
dd/mm/yyyy 12h, dd/mm/yyyy 24h, mm/dd/yyyy 12h, mm/dd/yyyy 24h, yyyy/mm/dd
12h, yyyy/mm/dd 24h.
Date, time zone and time are synchronized to UTC (coordinated universal time) with
desktop software only. This synchronization is recommended for the most accurate
datalog time records between the logger and the desktop. There is no automatic
adjustment for daylight savings. The user must change the time manually.
Sets the time zone (default: MST- US/Canada).
Sends or receives data or program files from the USB ports. Refer to
the instrument settings
on page 38 for more information.
Adjusts the brightness of the display (default: 50%).
Enables security for password protection. When enabled, the user must enter the
password to change the settings.
When the display goes to sleep or the instrument power is set to off, the user must
enter the password again. For password recovery, contact flow technical support.
Sets the display language.
Unit System—Sets the system of measurement that shows on the display. After the
unit system is selected, only the temperature units can change in the datalog
programming menu. Options: US Customary or Metric. Select Units—Sets the
individual measurement units that show on the display (e.g., level, velocity, flow,
temperature, surface velocity, distance, minimum distance, maximum distance, surge
level, rain, level raw and surge velocity).
Erases the selected log(s). Options: Data Log, Event Log, Alarm Log, Diagnostics
Log, Sample Log and Clear All.
Sets all the controller settings back to the factory defaults. Erases all data logs.
on page 26.