Error Messages 
 
Meter will not switch ON 
 Press the power button firmly for at least 0.1 seconds 
 Check that the batteries are installed, fresh, and correctly oriented 
 Remove one battery for one minute to reset the circuit, replace and try again 
 
Display switches OFF 
 The battery voltage may have dropped below operational requirements. Check the 
batteries and replace if necessary 
 The automatic power OFF (sleep mode) function may have switched the meter OFF. 
Switch the meter ON and if it does not power up replace the batteries and try again. 
 
E2 Error 
The reading is lower than the low range limit. Test the meter in an environment known to be 
within acceptable meter limits. If error persists, send the unit to Extech for repair. 
 
E3 Error 
No field solution. Return the unit to Extech Instruments for repair. 
 
E4 Error 
The reading is higher than the high range limit. Test the meter in an environment known to be 
within acceptable meter limits. If error persists, send the unit to Extech for repair. 
 
E31 Error 
Temperature related circuit error. Return the meter to Extech Instruments for repair. 
 
E32 Error 
Memory IC error; reboot the meter and check it again; send it to Extech Instruments for repair 
if the error message persists. 
 
E33 Error 
RH measurement circuit failure; return the unit to Extech Instruments for repair. 
 
 
Shop for Data Logging products online at:
15
DataLoggerStore
www.
AN340-EU-EN V1.3 4/13
1.877.766.5412
.com

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AN340 and is the answer not in the manual?

Questions and answers

Table of Contents