Table 2. Guidance and manufacturer's declaration—
electromagnetic immunity—for all equipment and systems
IEC 60601 test
Immunity test
level
Electrostatic
±6 kV contact
discharge
(ESD) (IEC
±8 kV air
61000-4-2)
Electrical fast
±2 kV for
transient/
power supply
burst (IEC
lines
61000-4-4)
±1 kV for
input/output
lines
Surge
±1 kV
(IEC
differential
61000-4-5)
mode
±2 kV
common
mode
66
Communicator Patient Manual
®
LATITUDE
Electromagnetic
Compliance level
environment—guidance
±8 kV contact
The relative humidity
should be at least 15
±15 kV air
percent.
±2 kV for
Mains power quality
power supply
should be that of a
lines
typical commercial or
hospital environment.
±1 kV for
input/output
lines
±1 kV
Mains power quality
differential
should be that of a
mode
typical commercial or
hospital environment.
±2 kV
common
mode