14. Error Messages and Fault Clearance

Messages are displayed in plaintext and are self-explanatory.
Error Message /
Error
"Dry Strip"
"Wrong Strip"
"Wrong Position" The strip hasn't been pushed
"No Paper"
"Battery Low"
"Instrument
doesn't start"
In case a fault cannot be cleared by the aid of the instructions above, please
contact your local distributor or the MACHEREY-NAGEL Service.
96
Relax
®
Cause
The test strip wasn't dipped
completely
A wrong test strip has been
detected (wrong type)
into the strip retainer far
enough
Paper roll empty or printer flap
open
Batteries are low
Power supply not installed or
defect
Manual URYXXON
Solution
Repeat measurement with
a new strip
Use correct test strips
New measurement, place
strip in right position
Replace paper and close
printer flap
Exchange batteries or use
power pack
Check whether all connec-
tions are plugged in and
whether the power socket
is functioning
Relax EN V 3.00 / 03.16
®