8. Quality control testing

Quality control measurements should be performed regularly with check solu-
tions in order to ensure the correct functioning of the combination of equipment
and test strips.
Quality inspection tests are recommended:
• at the beginning of a day
• when introducing a new test strip LOT
• in case of doubtful test results
• when a different person is operating the equipment
Upon pressing
?
(Display 19 + Display 20).
MENU
Memory
?
Check mode
Settings
Display 19: Main menu

8.1. How to review old QC measurements

The equipment saves the results of the last 20 QC measurements in a separate
memory. They may be displayed by pressing
printed for documentation purposes.
78
Relax
®
on the start menu the "Check mode" screen will appear
Manual URYXXON
CHECK MODE
Memory
Measurement
Display 20: Check mode
(Display 20) and may be
Relax EN V 3.00 / 03.16
®