Table 2: Immunity (Stimulation mode) table according to IEC 60601-1-2, 3
Guidance and manufacturer's declaration – electromagnetic immunity
The ARYS™ me is intended for use in the electromagnetic environment specified below. The customer or user of
ARYS™ me should ensure that it is used in such an environment.
Immunity test
IEC 60601
standard
test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrostatic fast
± 2 kV for power
transient / burst
supply lines
IEC 61000-4-4
± 1 kV for
input/output lines
Surge
± 1 kV line to line
IEC 61000-4-5
± 15 kV line to earth
Voltage dips, short
<5 % U
interruptions and
40 % U
voltage variations on
70 % U
power supply lines
<5 % U
IEC 61000-4-11
Note: U
is the AC mains voltage prior to application of the test level.
T
Power frequency
3 A /m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance level
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
(0.5 cycle)
Not applicable
T
(5 cycles)
T
(25 cycles)
T
for 5 s
T
Not applicable
| table 2 electromagnetic immunity |
edition
rd
Electromagnetic environment – guidance
Excessive electrostatic discharge may lead
to an error on the tracker. Should this
happen, please proceed according to the
troubleshooting table on page 53.
–
–
Not applicable according to IEC 60601-1-2
4
ed. for non-magnetic field sensitive
th
devices
|
61
|