Table 2: Immunity (Stimulation mode) table according to IEC 60601-1-2, 3
Guidance and manufacturer's declaration – electromagnetic immunity
The ARYS™ me is intended for use in the electromagnetic environment specified below. The customer or user of
ARYS™ me should ensure that it is used in such an environment.
Immunity test
IEC 60601
standard
test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrostatic fast
± 2 kV for power
transient / burst
supply lines
IEC 61000-4-4
± 1 kV for
input/output lines
Surge
± 1 kV line to line
IEC 61000-4-5
± 15 kV line to earth
Voltage dips, short
<5 % U
interruptions and
40 % U
voltage variations on
70 % U
power supply lines
<5 % U
IEC 61000-4-11
Note: U
is the AC mains voltage prior to application of the test level.
T
Power frequency
3 A /m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance level
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
(0.5 cycle)
Not applicable
T
(5 cycles)
T
(25 cycles)
T
for 5 s
T
Not applicable

| table 2 electromagnetic immunity |

edition
rd
Electromagnetic environment – guidance
Excessive electrostatic discharge may lead
to an error on the tracker. Should this
happen, please proceed according to the
troubleshooting table on page 53.
Not applicable according to IEC 60601-1-2
4
ed. for non-magnetic field sensitive
th
devices
|