Applications; Laboratory Glassware (Lg) - Miele G 7836 CD Operating And Installation Instructions

Laboratory glassware washer
Hide thumbs Also See for G 7836 CD:
Table of Contents

Advertisement

Applications

Laboratory glassware (LG)

Wide necked glassware, wide necked
Erlenmeyer flasks and petri dishes or
glassware with a cylindrical form,
e.g. test tubes can be cleaned inside
and out with the rotating spray arm.
General glassware used in laboratories,
e.g. beakers, conical flasks, petri
dishes, test tubes, etc. can be
arranged in full size, half or quarter
inserts at the higher or lower level.
Mobile injector units are available for
direct injection spraying of narrow
necked flasks, pipettes, etc.
The following instructions relate only to
basic preparation and loading of
glassware.
Loading the machine
Removing excess soils
^ Empty all glassware before loading
into the machine.
,
Ensure that no acid or solvent
residues, especially hydrochloric
acid or chlorides get into the wash
cabinet.
^ Remove all agar residues from petri
dishes.
^ Remove blood clots and residues
from test tubes, etc.
^ Remove all stoppers, corks, labels,
sealing wax residues, etc.
^ Small parts - such as stoppers,
faucets and taps - should be secured
in suitable basket inserts.
22
Note
^ Petri dishes and the similar should be
placed in the correct insert with the
soiled side facing the center.
^ Insert pipettes with the pointed end
downwards.
^ Quarter inserts should be placed in
the mobile unit with at least 1
(3 cm) between them and the edge
of the mobile unit.
E 350 Injector unit 1/1
(E 380 with connection to TA)
Sufficient for narrow neck laboratory
glassware with 32 (33) jets/clips.
1
/
"
8