Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The FreeStyle and FreeStyle 5 are intended for use in the electromagnetic environment specified below. The
customer or the user of the FreeStyle and FreeStyle 5 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic
environment –
±6 kV contact
guidance
±8 kV air
IEC 61000-4-2
±2 kV for power supply
lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/output
lines
±1 kV line(s) to line(s)
Surge
IEC 61000-4-5
±2 kV line(s) to earth
<5 % U
(>95 % dip in U
for 0,5 cycle
40 % U
Voltage
(60 % dip in U
dips, short
interruptions
for 5 cycles
and voltage
variations on
70 % U
power supply
input lines IEC
(30 % dip in U
61000-4-11
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A / m
magnetic field
IEC 61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
28 — MN218-C4 A English
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
N/A
±1 kV line(s) to line(s)
<5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
T
for 5 sec
3 A / m
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
T
)
T
Mains power quality should be that
T
of a typical commercial or hospital
)
environment. If the user of the FreeStyle
T
and FreeStyle 5 require continued
operation during power mains
interruptions, it is recommended that
T
the FreeStyle and FreeStyle are powered
)
from an uninterruptible power supply
T
or a battery.
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.