English
Setting up several wireless groups ensures the reliable
transmission of measured values with the correct
address when using more than one examination room
each with similar equipment.
The maximum distance between transmitters and
receivers is approx. 10 metres. This range may be
reduced under certain local conditions, e.g. thickness
0
and type of wall partitions.
1
The following combination of devices is possible for
each wireless group:
• 1 baby scale
• 1 personal scale
• 1 length measuring rod
• 1 seca wireless printer
• 1 PC with seca USB wireless adapter

Channels

Within each wireless group, the devices communicate
with each other via three channels (C1, C2, C3). This
ensures reliable, troublefree data transmission.
When you set up a wireless group with this scale, the
device will suggest three channels guaranteeing opti-
mum data transmission. We recommend accepting the
channel numbers suggested.
C1
You can also select the channel numbers (0 to 99)
C2
C3
manually - for example if you want to set up more than
0
one wireless group.
The channels must be sufficiently far apart to ensure
trouble-free data transmission. We recommend a
spacing of at least 30 between channel numbers. Each
channel number may only be used for one channel.
Example configuration: channel numbers when setting
up 3 wireless groups within one surgery:
• wireless group 0: C1=_0, C2=30, C3=60
• wireless group 1: C1=10, C2=40, C3=70
• wireless group 2: C1=20, C2=50, C3=80
The seca 360° wireless network • 67