5-23. Sequential Point Plotting Time; Deflection Factor Test Setup - HP 1317A Operating And Service Manual

Table of Contents

Advertisement

Model13l7A
VOLTMETER CALIBRATOR
. .
-
*
-
I
1\
@
@
DOUBLE BANANA TO
BNC ADAPTOR
• •
{)l
50[2
COAX
X
INPUT
(IN REAR)
Model 1317A
MEASURE
LENGTH
~
Performance
Check
1317 A-L-008
Figure 5-4. Deflection Factor Test Setup
k. Set pulse generator delay to 10 ns and vernier
cwo Note clean bright dot on lower left of CRT.
1. Turn delay vernier ccw. Note appearance of
distorted trace segment projecting from dot.
m. Adjust delay vernier until length of trace
segment is one dot
diameter.
n. Measure time difference between two tran-
sitions displayed on monitor oscilloscope. Difference
shall be equal to or less than 500 ns.
5-23. SEQUENTIAL POINT PLOTTING TIME.
5-24. Required Result. The spot
shall
settle to
within
0.010
inch of final value in less than 200
nanoseconds for any
0.10
inch step.
5-25. Procedure.
a.
Perform
preoperational procedure
(Section
III).
b. Set
500/lOK
switches (on boards AI, A3, and
A5) to 500.
c. Connect equipment
as
shown
in
figure 5-5.
Note that line to Z INPUT is not yet connected.
d. Set square wave generator for 100-kHz sym-
metrical output.
e. Set amplitude of square wave generator to
cause 0.1 inch deflection on CRT.
f. Set BRIGHTNESS ccw.
g. Connect
Z
INPUT.
h. Set pulse
generator
for
100-kHz, IV
p-p,
llJ.sec
width pulse
output.
i.
Set sweep time of monitor oscilloscope
to
O.llJ.sec/div and display switch to ALT
B.
j. Set monitor oscilloscope trigger controls
to
achieve stable display. If necessary, slightly read-
just symmetry control on square wave generator to
achieve synchronization between channels A and B.
k. Set pulse generator delay to 10 ns and ver-
nier cwo Note single clean bright dot.
5-5

Advertisement

Table of Contents
loading

Table of Contents