Deflection Amplifier Phase Shift; Risetime Test Setup - HP 1317A Operating And Service Manual

Table of Contents

Advertisement

Performance Check
MONITOR OSCI LLOSCOPE
0
'
/
(J
@
1 1-t-+-t-+--t---1r--;~-I
1
C)
I
f-- --+-+-+----+--'~-+-t-H
@
o
@
MEASURE
RISETIME
10:1 DIVIDER PROBE
PROBES CONNECTED TO HORIZONTAL
DEFLECTION PLATES (NECK PINS ON CRT).
SQUARE WAVE GENERATOR
I
_ _
0
1
....
son
COAX
@
@)
e e
son
OUTPUT
X
INPUT
(IN REAR)
Model 1317A
Model 1317A
1317 A-L-005
Figure 5-1. Risetime Test Setup
e. Set amplitude of constant amplitude signal
generator for 5-inch display on CRT.
f.
Using POSITION X and POSITION Y con-
trols, center display on CRT.
g. Change frequency of constant amplitude
signal generator
to
5 MHz. CRT display shall be equal
to
or greater than 3.7 inches.
h
.
Disconnect
constant
amplitude
signal
generator from X INPUT and connect
to
Y INPUT.
i. Repeat steps d through g.
5-14.
DEFLECTION AMPLIFIER PHASE SHIFT.
5-15. Required Result. Phase shift between the
x and y amplifier shall be less than 0.1 degree
to
50
kHz and less than 1 degree to 250 kHz for full screen
signals.
5-2
5-16. Procedure.
a. Perform preoperational procedure (Section
III).
b. Set
500/10K
switches (on boards Al and A3)
to 500.
c. Connect equipment as shown in figure 5-3.
d. Set frequency of constant amplitude signal
generator
to
50 kHz.
e. Set amplitude of constant amplitude signal
generator for full screen display (corner
to
corner)
on CRT.
f. Using POSITION X and POSITION Y, center
display. There shall be no noticeable separation of
diagonal trace.