HP 1317A Operating And Service Manual page 9

Table of Contents

Advertisement

Model 1317A
General Information
Table 1-1. Specifications (Cont'd)
SAFETY PROTECTION
IMPLOSION:
exceeds safety requirements of
IEC 348 (IEC 65) and ANSI C39.5 for Electron-
ic Measuring Apparatus.
HIGH VOLTAGE:
anode lead is permanently
bonded to CRT, minimizing shock hazard. All
high potentials are completely enclosed or
covered in compliance with UL recommenda-
tions.
X-RAY EMISSION:
<0.1 mr/hr. Measured with
Victoreen Model 440 RF/C in background
noise.
GENERAL
X, Y, AND Z INPUT CONNECTORS:
BNC type
mounted to rear panel.
DIMENSIONS:
See outline drawing.
FRONT PANEL CONTROLS:
Brightness, Posi-
tion
X,
Gain
X,
Position
Y,
Gain
Y,
Trace
Align, Orthogonality,
Focus,
and Astigmatism
T
q
.-
~
v
I
M
located below CRT and hidden behind a con-
trol door.
LINE
INDICATOR:
lamps
mounted
behind
front panel door and on rear panel.
POWER:
selectable to 100 Vac, 120 Vac, 220 Vac
and 240 Vac +5 % -10%, 48 Hz to 440 Hz,
100 VA maximum.
ENVIRONMENT
TEMPERATURE:
-40°C to +70°C non-operating.
O°C to +50°C operating.
HUMIDITY:
50-95% relative
humidity
from
25°C to 40°C. (per MIlrE-16400E, class
3 and
4.)
ALTITUDE:
25,000 feet non-operating. 15,000
feet operating.
VIBRATION:
vibrated in three planes for 15
minutes each with 0.010 to 0.060 peak
to
peak
excursion. 10-55 Hz (per MIlrSTD-167).
SHOCK:
level 30G; duration 11 ms; shape
1/2 sinewave (Reference MIlrE-4970A, Pro-
cedure II).
o
o
o
o
o
o
o
f
~-------l~
i
lJ__
j~
o
o
[
C - 1 6 3/4 (425,5)
~-
~~----211/4(539'8)-----I
..
~1
'-'7/8(22,2)
.... - - - - 225/16 (566,7)
-------I-:-~
SIDE
VIEW
19
(482,6)----~
REAR
VIEW