Setup; Required Equipment - Fluke 1732 Calibration Manual

3540 fc logger/monitor
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

Setup

Before you start the verification procedures or make calibration adjustments, refer to this section for the
equipment, system, and setup requirements.

Required Equipment

See Table 3 for a list of requirements for the verification tests and calibration adjustment of the Logger.
Calibrator
Cable Assembly
173x AUX Input Calibration
Cable
[1]
173x Calibration Cables –
Voltage-to-Current Input Cable
Assembly
173x Verification Box
USB cable
DMM
Coil
Coil
Banana-to-Pin Adapter
[1] The 173x calibration cables and verification box are not available from Fluke. See Equipment Assembly for
information on how to make these items.
Equipment
5522A
3PHVL-1730
NA
NA
[1]
[1]
NA
type A-to-mini B
8846A
5500A/COIL
Optional: 52120A
with Coils
NA
Pomona
Electronics 4690
Table 3. Required Equipment
Model
Notes
5520A is also supported
Voltage Test Lead
3-Phase+N
1732/1734/1736/1738 only
1732/1734/3540 FC:
Qty. 3 required
1736/1738:
Qty. 4 required
for AUX Adapter
verification
for Flexi verification
5 turns
for AUX Adapter
verification
Logger/Monitor
Setup
Used on:
Verification
Calibration
Tests
Adjustment
X
X
X
X
X
X
X
X
X
optional
X
X
X
X
X
X