173x AUX Input Calibration Cable

See Table 5 and Figure 1 for instructions on how to make the calibration cable.
Cable must be marked with "max. 30 V to earth." Any voltage-, category-, or
current-ratings on safety plugs must be removed.
Item
Table 5. 173x AUX Input Calibration Cable
Description
Binder: Series 620 - Male Cordset, 4-pole, 2 m
Test Lead 0.75 mm² with 4 mm Banana Plug, stackable
Test Lead 0.75 mm² with 4 mm Banana Plug, stackable
Shrink tube Ø 5-6 mm, black, thin wall, 3:1
Shrink tube Ø 8-10 mm, transparent, thin wall, 2:1
Shrink tube Ø 10-12 mm, black, thin wall, adhesive, 3:1
Shrink tube Ø 12-14 mm, black, thin wall, 3:1
Caution
Binder: 79 9241 020 04
red
black
L = 30 mm
L = 45 mm
L = 30135
L = 110 mm
Logger/Monitor
Part Number/Info
1
2
2
4
2
1
1
Setup