Hanna Instruments HI 5421 Instruction Manual

Hanna Instruments HI 5421 Instruction Manual

Do/bod/our/sour/temperature bench meter
Hide thumbs Also See for HI 5421:

Advertisement

Instruction Manual
HI 5421
Hanna Instruments Inc.
Highland Industrial Park
DO/BOD/OUR/SOUR/Temperature
584 Park East Drive
Woonsocket, RI 02895 USA
Bench Meter
Technical Support for Customers
Tel. (800) 426 6287
Fax (401) 765 7575
E-mail tech@hannainst.com
www.hannainst.com
Local Sales and Customer Service Office
MAN5421 03/15
Printed in ROMANIA
w w w. h a n n a i n s t . c o m

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 5421 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 5421

  • Page 1 Instruction Manual HI 5421 Hanna Instruments Inc. Highland Industrial Park DO/BOD/OUR/SOUR/Temperature 584 Park East Drive Woonsocket, RI 02895 USA Bench Meter Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.com Local Sales and Customer Service Office...
  • Page 2: Table Of Contents

    If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service Department and then send it with shipping costs prepaid.
  • Page 3: Preliminary Examination

    . s g HI 5421 is a professional bench meter with color graphic LCD for DO, BOD, OUR, SOUR and temperature r t s n t n e e o d...
  • Page 4: Functional Description

    FUNCTIONAL DESCRIPTION PROBE CONDITIONING & MAINTENANCE The DO probe body is made of PEI. HI 5421 DESCRIPTION Use the protective cap provided when the probe is not in use. To inspect or replace the FRONT PANEL membrane proceed as follows: Inspection: Inspect membrane surface to ensure it is in perfect condition.
  • Page 5 To enter / exit calibration mode. software (optional). HI 92000 also offers graphing and on-line help features. Data logged on the HI 5421 meter can be exported to the most popular spreadsheet applications for further To select the desired measurement mode: DO, BOD, OUR, SOUR.
  • Page 6: Specifications

    To delete lots: SPECIFICATIONS • Press while in Log Recall mode. e g n 0 . 0 0 . 0 0 . 0 0 . 0 u t a t a r n o i • Press to access delete or delete all o s e t u l n o i...
  • Page 7: Operational Guide

    • Press to select the desired Log OPERATIONAL GUIDE Report type. All logged lots for the selected Log Report type will be displayed on the LCD. POWER CONNECTION • To filter the displayed lots, press and then the Plug the 12 Vdc adapter into the power supply socket. desired parameter.
  • Page 8: Displaying Modes

    3) Direct/AutoHold Reading Mode and Automatic Log DISPLAYING MODES Press and then keys must be pressed on front For each measurement mode (DO, BOD, OUR or SOUR) the following display configurations are available: display to initiate this function. Real time continuous measurements Basic, Graph and Log History.
  • Page 9 Graph LOGGING The online graph with real time logging (DO, BOD, OUR or SOUR There are 5 ways the Reading Mode and Log may be configured together. The table below shows the vs. Seconds) will be displayed when this option is selected. combinations and indicates where the completed log will be stored.
  • Page 10 Log History SOUR MEASUREMENT The measurement, along with LOG History, will be visible when this option is selected: The Specific Oxygen Uptake Rate (SOUR), also known as the oxygen consumption or respiration rate, is defined as the milligram of oxygen consumed per gram of volatile suspended solids (VSS) per hour. This quick measurement 1) The last stored logged data (Not actively logging) or has many advantages: rapid measure of influent organic load and biodegradability, indication of the presence of 2) The last data logged from an active logging lot or...
  • Page 11: System Setup

    OUR MEASUREMENT SYSTEM SETUP The OUR is used to determine the oxygen consumption or respiration rate. It is defined as the mg/L of oxygen The System Setup menu allows the user to customize the user interface, view meter information, set the consumed per hour.
  • Page 12 To print a copy of the BOD analysis with seed correction: desired option. • Use HI 92000 software and connect the HI 5421 meter. • Press to edit the desired information. The Text Editor • Select the BOD sample from the available BOD reports.
  • Page 13 SEED CORRECTION Date & Time Set the current date & time and the format in which they appear. These parameters will be displayed on the To apply seed corrections to the sample data follow the following Measure screens and also when storing measured data. procedure: Set Date and Time •...
  • Page 14 • Use HI 92000 software to connect the HI 5421 meter. • Select BOD sample fom available BOD samples. • Press and use to highlight the desired color.
  • Page 15 Language • With the probe measuring the sample press . The initial This option allows the user to choose the desired language in which all informations will be displayed. DO measurement will be saved. • Remove probe from sample and fill and cover to prepare bottle To select the Language: for incubation.
  • Page 16 To view the Meter Information: BOD MEASUREMENT • Press while in Measure mode. Biochemical oxygen demand (BOD) is an indicator for the concentration of biodegradable organic matter • Press present in a sample of water. It can be used to determine the general quality of the water and its degree •...
  • Page 17: Do Setup

    The meter contains a built-in barometer, and it is able to automatically compensate for changes in barometric DO SETUP pressure. If another pressure value than the barometer’s reading is to be used, then the manual pressure feature must be enabled (See DO Setup ). The DO Setup menu allows the user to set the parameters related to the DO measurement and calibration.
  • Page 18 Save Profile BAROMETRIC PRESSURE COMPENSATION To save a profile: The dissolved oxygen saturation value varies with pressure, so it is important to compensate the effect that • Press while in DO mode. pressure has on DO measurements. • Press i t l C °...
  • Page 19 SALINITY COMPENSATION • Press . A list with all customized profiles will be displayed on the screen. If the sample contains significant concentration of salinity, the read out values must be corrected, taking into • Use to select the desired profile and press to confirm or to exit without account the lower degree of oxygen solubility in this situation.
  • Page 20: Do Measurement

    Temperature Unit DO MEASUREMENT The user can choose from Celsius , Fahrenheit or Kelvin temperature units. Make sure the instrument has been calibrated before taking DO measurements. To set the Temperature Unit : • Press while in DO mode. DIRECT MEASUREMENT •...
  • Page 21 Calibration Reminder When the user standard is selected: This option allows the user to set the calibration reminder as Daily , Periodic or Disabled . • The calibration can be performed at one value. To set the calibration reminder: • Edit the desired standard value by using then •...
  • Page 22 Clear Calibration Note: When not in use and during polarization, use the protective transparent cap. Accessing this option, the existent DO calibration can be cleared. If the calibration is cleared a default meter calibration is used. Another sensor calibration should be performed. To clear calibration: •...
  • Page 23 Barometer DO CALIBRATION This parameter allows configuration of parameters related to barometric pressure. This parameter is only needed when measuring in concentration units of mg/L or ppm. It is recommended to calibrate the probe frequently, especially if high accuracy is required. From the Barometer menu the user can choose the pressure source and units, as well as the pressure.
  • Page 24 Pressure Units SOUR SETUP The user can choose from six pressure units: mmHg , mbar , kPa , The SOUR Setup menu allows the user to set the parameters related to SOUR measurement. inHg , psi or atm . To set the Pressure Unit: Accessing SOUR Setup •...
  • Page 25 • Use to highlight the Salinity Unit option. OUR SETUP to select the desired units . • Press and then use The OUR Setup menu allows the user to set the parameters related to the OUR measurement. • Press to confirm your selection or press to cancel operation.
  • Page 26: Bod Measurement

    • For text editing use to highlight the BOD SETUP desired character and then press to add it to the text bar. It is also possible to delete the last character by positioning The BOD Setup menu allows the user to set the parameters related to the BOD measurements. the cursor on the Backspace character ( ) and pressing Accessing BOD Setup •...
  • Page 27: Logging

    • Press and use highlight the Alarm State option. Note: See Logging section for available types of logging. • Press and use to highlight the desired option. This option allows the user to edit the log settings: Logging Type , Logging Data Configuration , Sampling Period •...
  • Page 28 New Lot • Press while in DO mode. This option is used to create a new lot when manual logging is used. • Press Note: If New Lot option is accessed and the Logging Type is • Use to highlight the Log option. Automatic , a warning message appears on the LCD informing •...

Table of Contents