Em50 Series
stored data.
4.2.5

Instantaneous Measurements (Scan)

ECH20 Utility gives you the ability to take real-time sensor measure-
ments with your Em50 logger. This is most useful as a troubleshoot-
ing feature to test if a sensor is reading properly. You can also see
how sensors react to environmental changes. Follow steps 1 and 2 to
take real time measurements.
1. Click the "Scan" button in the toolbar, or go to Actions
Logger Ports. The Em50 takes a reading on each of the ports,
then displays the Em50 Port Scan screen.
2. The Port Scan screen displays all five ports should display with
the measurement for each port in the units appropriate for the
sensor. You can change measurement units in the Preferences
Menu (Section 4.3.2). Click the sensor name to see the reading
as raw data.
The five ports should display with the measurement for each port
in the current selected unit. If nothing is plugged in to a port, the
reading for that port is zero.