Selecting the Temperature Unit of Measure (°C/°F)
Pull the trigger and release. Use the M button to step to °C/°F. Use the SET button to select °C or
°F. Pull the trigger to return to the normal operating mode.
Setting the Alarm Buzzer 'bU2' ON/OFF
Pull the trigger and release. Use the M button to step to 'bU2'. Use the SET button to select ON
or OFF. Pull the trigger to return to the normal operating mode. When set to OFF, the Alarm
buzzer will not sound in an alarm condition.
Measurement Notes
1. The object under test should be larger than the spot (target) size calculated using the field of
view diagram.
2. If the surface of the object under test is covered with frost, oil, grime, etc., clean before
taking measurements.
3. If an object's surface is highly reflective apply masking tape or flat black paint to the surface
before measuring.
4. The meter cannot make accurate measurements through transparent surfaces such as glass.
5. Steam, dust, smoke, etc. can obscure measurements.
6. The meter compensates for deviations in ambient temperature. It can, however, take up to
30 minutes for the meter to adjust to extremely wide ambient temperature changes.
7. To find a hot spot, aim the meter outside the area of interest then scan across (in an up and
down motion) until the hot spot is located.