Specifications
Temperature Specifications
Temperature Mode
Range
Accuracy
Resolution
Emissivity
Field of View
Laser pointer
IR Spectral
response
* The supplied thermocouple is rated for 250
temperatures, please obtain a Type‐K thermocouple rated for the desired temperature.
General Specifications
Display
Display update rate
Operating Temperature
Operating Humidity
Storage Temperature
Power Supply
Automatic Power Off
Weight
Dimensions
.
800.561.8187
Object (Surface Mode)
‐50 ~ 600°C (‐58 ~ 1112°F)
± 5°C (4.1°F) from ‐50°C ~ ‐15°C (‐58 ~ 5°F)
4°C (± 3.9°F) from ‐15°C ~ 0°C (5 ~ 32°F)
2°C (± 3.6°F) from 0°C ~ 100°C (32 ~ 212°F)
± 3%rdg + 1°C (1.8°F) from 100°C ~ 600°C (212 ~ 1112°F)
Accuracy is specified for the following ambient temperature range: 23 ~ 25°C (73 ~77°F)
0.1 °C/°F
0.05 ~ 1.00 adjustable
D/S = Approx. 12:1 ratio (D = distance; S = spot or target)
Class 2 laser < 1mW power; Wavelength is 630 ~ 670nm
8 ~ 14 mm (wavelength)
Backlit multifunction LCD display
< 1 second approx.
o
0
C ~ 50
10 ~ 80% Relative Humidity max.
o
‐10
C ~ 60
One 9V battery
Meter shuts off automatically after 10 seconds
138g (4.88 oz.)
153 x 120 x 42mm (6.0 x 4.7 x 1.7")
Copyright © 2016 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
www.
Ambient (Air)
‐20 ~ 70°C (‐4 ~ 158°F)
o
o
C (482
F) maximum temperature. To measure higher
o
o
o
C (32
F ~ 122
F)
o
C (‐14 ~ 140°F)
11
.com
Thermocouple (Type K)
‐50 ~ 1000°C (‐58 ~ 1832°F)*
IR267‐en‐GB_v1.1 8/16
information@itm.com
Need help?
Do you have a question about the IR267 and is the answer not in the manual?
Questions and answers