Chapter 2 – Important Safety Information
2.9
Verification of Proper
Functionality
A power supply (AC adapter with mi-
cro-USB connector) is provided as part
of the system. Before using your pump,
ensure that the following occur when
you connect a power supply into the
USB port of your t:slim X2 Pump:
You hear an audible alert
•
Your see the green light illuminate
•
from the edge around the Screen
On/Quick Bolus Button
You feel a vibratory alert
•
You see a charge symbol (lightning
•
bolt) on the battery level indicator
In addition, before using the pump,
ensure the following:
Press the Screen On/Quick Bolus
•
Button to turn the screen on so that
you can see the display
When the display screen is on, the
•
touch screen responds to your
finger tap
30
PRECAUTION
CONFIRM that the screen display turns on, you
can hear audible beeps, feel the pump vibrate,
and see the green LED light blinking around
the edge of the Screen On Button when you
connect a power source to the USB port. These
features are used to notify you about alerts,
alarms, and other conditions that require your
attention. If these features are not working,
discontinue use of the pump and contact
Tandem Diabetes Care Customer Technical
Support at (877) 801-6901.