Signet 8550-1 Installation Manual page 5

Flow transmitter
Hide thumbs Also See for 8550-1:
Table of Contents

Advertisement

Step 1.
Press and hold ENTER key:
Step 2.
The Key Code is UP-UP-UP-DOWN keys in sequence.
Step 3.
Scroll menu with UP or DOWN arrow keys.
Step 4.
Press RIGHT ARROW key to select menu item to be edited.
Step 5.
Press UP or DOWN keys to edit the flashing element.
Step 6.
Press ENTER key to save the new setting and return to Step 3.
Notes on Step 1:
• The View Menu is normally displayed.
• The CALIBRATE and OPTIONS menus require a KEY CODE.
VIEW
CALIBRATE
OPTIONS
Notes on Steps 3 and 4:
• Refer to pages 6 and 7 for complete listing of menu items and their use.
• From the Step 3 display, pressing the UP and DOWN keys simultaneously will
return the display to the VIEW menu.
• If no key is pressed for 10 minutes, display will also return to the VIEW menu.
Step 3: Finished Editing?
Press the UP and DOWN keys simultaneously after
saving the last setting to return to normal operation.
Notes on Steps 5 and 6:
• All output functions remain active during editing.
• Only the flashing element can be edited.
• RIGHT ARROW key advances the flashing element in a continuous loop.
• Edited value is effective immediately after pressing ENTER key.
• If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.
• Step 6 (pressing ENTER key) always returns you to Step 3.
• Repeat steps 3-6 until all editing is completed.
Step 5: Made an Error?
Press the UP and DOWN keys simultaneously
while any element is flashing. This will recall the
last saved value of the item being edited and
return you to Step 3.
‡ SIGNET 8550-1 Flow Transmitter
ProcessPro Editing Procedure:
• 2 seconds to select the CALIBRATE menu
• 5 seconds to select the OPTIONS menu.
• After entering the Key Code, the display will show the first item in the selected menu.
• The first display element will begin flashing.
• RIGHT ARROW key advances the flashing element.
Press &
ENTER
hold for
access:
2s
5s
Notes on Step 2:
If no key is pressed for 5 minutes while display is showing "Enter
Key Code", the display will return to the VIEW menu.
CALIBRATE: ----
Enter Key Code
CALIBRATE: *---
Enter Key Code
CALIBRATE: **--
Enter Key Code
CALIBRATE: ***-
Enter Key Code
Flow Units:
GPM
Flow Units:
GPM
Step 3
Output Setpnt:
20.0 GPM
Output Setpnt:
2
0.00 GPM
First item in
>
CALIBRATE menu
>
Step 4
>
Output Setpnt:
Step 5
1
0.00 GPM
Output Setpnt:
0
1
.00
GPM
Output Setpnt:
9
1
.00 GPM
ENTER
Step 6
Output Setpnt:
Saving
Output Setpnt:
19.00 GPM >
page 5 of 8