3-7 Digital calibration

Set CAL switch to ON
Select Digital calibration
Press the
Two sec. later
Method 1
q
Input zero voltage
Press the
Two sec. later
Input the span voltage
Press the
Two sec. later
Enter the weighing capacity
Press the
Display shows nothing
after calibration
Calibration completed set calibration switch to the OFF position
Please refer to the error message list if the display shows
04011210
Key
.
Method 2
With no weight on the platform or in the hopper
press the
.
key
.
.
key
.
key
Example:
mV/V
Zero voltage
(ZERO)
Zero Voltage ⇒ 0.00036 mV/V(incl. dead load)
Span Voltage ⇒ 2.90000 mV/V
Capacity ⇒ 30 000 divisions
key to set zero.
22
Span voltage
(SPAN)
Capacity
( CAP. )
Weight
⇒ Increment flashing digit
⇒ Decrement flashing digit
⇒ Move flashing point left
⇒ Move flashing point right
⇒ Store data in memory
⇒ Exit
Escape
/
. X
.
ZSME400000003