Excell EX2005 User Manual page 51

Table of Contents

Advertisement

EXCELL PRECISION CO., LTD
Function setting
Press the
key
Input desired
Parameter code
Press the
key
Display shows the existing
parameter code. Input
new code as required.
Press the
key
Continue to another function setting
or press
Item
Function
AnL- 01
Data type
AnL- 02
Signal output
AnL- 03
Weight in Lo
Current / Voltage
AnL- 04
in Lo
AnL- 05
Weight in Hi
Current / Voltage
AnL- 06
in Hi
04011210
key to escape
Parameter
0
1
2
0
1
When the weight reaches the value of
000000 ~ 999999
that in AnL-03(the Max. weight is the
max setting value (refer to CSP-04 ),
0.0 mA ~ 20.0 mA
the current / voltage(refer to AnL-02)
or
output is changed to that configured in
0.0 V ~ 10.0 V
When the weight reaches the value of
000000 ~ 999999
that in AnL-05(the Max. weight is the
max setting value (refer to CSP-04 ),
0.0 mA ~ 20.0 mA
the current / voltage(refer to AnL-02)
or
output is changed to that configured in
0.0 V ~ 10.0 V
49
⇒ Increment flashing digit
⇒ Decrement flashing digit
⇒ Move flashing point left
⇒ Move flashing point right
⇒ Store data in memory
⇒ Exit
Setting value
Description
As display
Gross
Net
Current output
Voltage output
AnL-04
AnL-06
Escape
/
Default
0
0
0
4.0
300000
20.0
ZSME400000003

Advertisement

Table of Contents
loading

Table of Contents