Table 29: Between-Instrument Precision
 
Percent of specimens that have three
matching NILM replicates or three
matching ASC-US+ replicates
Percent of specimens that have three
matching <LSIL replicates or three
matching LSIL+ replicates
Percent of specimens that have three
matching <HSIL replicates or three
matching HSIL+ replicates
Percent of specimens that have three
matching Satisfactory replicates or
three matching UNSAT replicates
* 120 specimens were enrolled, but 5 were excluded due to slide breakage and other errors.
 
 
Cell Count Study
 
The quantity of cellular material transferred onto slides, comparing ThinPrep 5000 to the ThinPrep
2000, was evaluated in a laboratory study using a split-sample technique.
Two hundred ten (210) specimens were enrolled in the study (139 NILM, 28 ASC-US, 28 LSIL, and
15 HSIL). Each specimen was split into two parts, processed on a ThinPrep 2000 and ThinPrep 5000
system, then stained and coverslipped. All slides were run on a ThinPrep Imaging System to obtain
Imager object count data, which has been demonstrated to correlate closely with cytotechnologist cell
count estimates. Cellularity varies among clinical specimens, so a range of cell counts was obtained.
The chart below provides a scatter plot of the count data from the matched pairs of slides in this
study. The Control axis is the ThinPrep 2000 slide's count value, and the Test axis is the matching
ThinPrep 5000 slide's count.
 
ThinPrep 5000
94.0%
(110/117)
(88.2% to 97.1%)
97.4%
(114/117)
(92.7% to 99.1%)
98.3%
(115/117)
(94.0% to 99.5%)
98.3%
(117/119)
(94.1% to 99.5%)
ThinPrep 2000*
91.1%
(102/112)
(84.3% to 95.1%)
94.6%
(106/112)
(88.8% to 97.5%)
100%
(112/112)
(96.7% to 100%)
98.3%
(113/115)
(93.9% to 99.5%)
MAN-04008-001 Rev. 001
Page 24 of 27