Agreement among Laboratory Cytotechnologists/Pathologists
The following tables indicate the extent to which the laboratory cytotechnologists / pathologists at a given site
agreed amongst themselves on the diagnosis, comparing the ThinPrep 5000 processor to the ThinPrep 2000
system. Tables are provided for ASC-US+ and ASC-H+.
In Table 24 for ASC-H+, the number of specimens is shown for which various levels of agreement among
the CTs occurred. Either all three CTs rated the slide as positive (ASC-H+), two out of three rated it
positive, one out of three, or none of them.
ThinPrep
Three CTs had
5000
Processor
Two CTs had ASC-H+
Three lab
and one had <ASC-H
CTs have
read the
One CT had ASC-H+
same
and two had <ASC-H
ThinPrep
Three CTs had
5000 slide
from a vial
 
 
 
 
 
 
 
ThinPrep 5000
 
Processor
 
 
Three lab CTs have
 
read the same
 
ThinPrep 5000 slide
 
from a vial
 
 
 
 
Table 24: Laboratory Cytotechnologist/Pathologist Agreement, All Results, ASC-H+
Three lab CTs have read the same ThinPrep 2000 slide from a vial
Three CTs
ASC-H+
had
ASC-H+
 
111
ASC-H+
 
32
 
7
 
2
<ASC-H
Totals
152
ASC-H+
 
Three or two CTs
had ASC-H+
 
Three or two CTs
had <ASC-H
Totals
ThinPrep 2000 System
Two CTs had ASC-H+
& one had <ASC-H
21
30
9
8
68
ThinPrep 2000 System
Three lab CTs have read the same
ThinPrep 2000 slide from a vial
Three or two CTs
had ASC-H+
194
26
220
One CT had ASC-H+
CTs had
& two had <ASC-H
<ASC-H
 
6
 
21
 
43
 
37
107
 
Three or two CTs
Totals
had <ASC-H
34
1006
1032
1040
MAN-04008-001 Rev. 001
Three
Totals
0
138
7
90
28
87
898
945
933
1260
242
1260
Page 20 of 27