Section 3: Basic DMM operation

Thermocouple measurements

The following table lists the temperature range for the supported thermocouple (TC) types:
Thermocouple temperature measurement ranges
Thermocouple type
B
C
E
J
K (default)
N
R
S
T
Configuration settings for thermocouple temperature measurements:
Thermocouple type: B, C, E, J, K, N, R, S, T
Default setting: K
Measurement units: °C or °F
Default setting: °C
Reference junction: Real or simulated
Default setting: Real
A thermocouple measurement is referenced to a known temperature. The reference junction can be real
or simulated.
For a real reference junction, the instrument performs a temperature measurement on a built-in cold
junction compensator and uses that reading as the reference point for subsequent thermocouple
measurements.
For a simulated reference junction, you enter a temperature reading to use as the reference
temperature. An ice bath is typically used in determining a simulated reference junction.
3-22
Temperature range (°C)
600 to 1820
0 to 2316
-250 to 1000
-210 to 1200
-200 to 1372
-200 to 1300
0 to 1767
0 to 1767
-250 to 400
Model 2110 5½ Digit Multimeter Reference Manual
Temperature range (°F)
1112 to 3308
32 to 4200.8
-418 to 1832
-346 to 2192
-328 to 2501.6
518 to 2372
32 to 3212.6
32 to 3212.6
-418 to 752
2110-901-01 Rev. C/August 2013