Performance Tests For The 15B+ And 17B - Fluke 15B+ Calibration Manual

Hide thumbs Also See for 15B+:
Table of Contents

Advertisement

Performance Tests for the 15B+ and 17B+

Before doing the performance test, make sure to warm up the calibrator.
To test each function and operating ranges:
1. For each step in Table 3, set the UUT to the specified function and range.
2. Connect the source to the input jacks on the UUT.
3. Apply the output from the source.
4. Make sure the readings on the UUT display are within the limits in Table 3.
Table 3. Performance Specifications for 15B+ and 17B+
Step
Function
1
2
3
VAC
4
5
6
7
VAC/Freq.
8
9
VAC/Duty
10
11
VDC
12
13
14
mVDC
15
mVAC
Calibrator output
Frequency
Value
or Amplitude
3.5 V
50 Hz
35 V
50 Hz
35 V
500 Hz
350 V
50 Hz
1000 V
50 Hz
45 Hz
1 V
5 KHz
1 V
100 KHz
3 V
3 V (Square
50 Hz
Wave)
3.5 V
0 Hz
35 V
0 Hz
350 V
0 Hz
-1000 V
0 Hz
350 mV
0 Hz
350 mV
500 Hz
UUT reading limit
15B+
17B+
Lower
Limit
3.462
34.62
34.62
346.2
986
X
44.92
X
4.992
X
99.6
X
49.5
3.480
34.80
348.0
-1008
345.4
339.2
Digital Multimeter
Performance Tests
Upper
Limit
3.538
35.38
35.38
353.8
1014
45.08
5.008
100.4
50.5
3.520
35.20
352.0
-992
354.6
360.8
11