Fluke 15B+ Calibration Manual page 19

Hide thumbs Also See for 15B+:
Table of Contents

Advertisement

15B+/17B+/18B+
Calibration Manual
Step
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
14
Table 4. Performance Specifications for 18B+ (cont.)
Function
Value
0 Ω
350 Ω
3.5 kΩ
OHM
35 kΩ
350 kΩ
3.5 mΩ
10 mΩ
40 Ω
BEEPER ON
DIODE
0.7 V
35 nF
350 nF
3.5 μF
CAP
35 μF
350 μF
3.5 A
ADC
-10 A
3.5 A
AAC
10 A
35 mA
mADC
-350 mA
35 mA
mAAC
350 mA
350 μA
μADC
-3500 μA
350 μA
μAAC
3500 μA
-5.5 V
LED
5.5 V
Calibrator Output
Frequency or
Amplitude
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
0 Hz
Ω
Ω
Ω
Ω
Ω
0 Hz
0 Hz
50 Hz
400 Hz
0 Hz
0 Hz
400 Hz
50 Hz
0 Hz
0 Hz
40 Hz
400 Hz
0 Hz
0 Hz
UUT reading limit
Lower Limit
Upper Limit
-0.3
0.3
348.0
352.0
3.480
3.520
34.80
35.20
348.0
352.0
3.480
3.520
9.82
10.18
X
X
0.622
0.778
32.75
37.25
342.4
357.6
3.320
3.680
33.20
36.80
332.0
368.0
3.444
3.556
-10.18
-9.82
3.444
3.556
9.82
10.18
34.44
35.56
-355.6
-344.4
34.44
35.56
344.4
355.6
344.4
355.6
-3556
-3444
344.4
355.6
3444
3556
-6.05
-4.95
4.95
6.05

Advertisement

Table of Contents
loading

This manual is also suitable for:

17b+18b+

Table of Contents