Omega CL540A User Manual page 4

Thermocouple source
Hide thumbs Also See for CL540A:

Advertisement

Ranges & Accuracies
T/C
Degrees C
Accuracy
Degrees F
Type
Range
J
-200.0 to -180.0
±0.5°
-346.0 to -292.0
-180.0 to -50.0
±0.4°
-292.0 to -58.0
-50.0 to 500.0
±0.3°
-58.0 to 932.0
500.0 to 1200,0
±0.4°
932.0 to 2192.0
K
-230.0 to -100.0
±0.8°
-382.0 to -148.0
-100.0 to 1050.0
±0.4°
-148.0 to 1922.0
1050.0 to 1371.1
±0.5°
1922.0 to 2500.0
T
-260.0 to -200.0
±1.2°
-436.0 to -328.0
-200.0 to -50.0
±0.7°
-328.0 to -58.0
-50.0 to 0.0
±0.4°
0.0 to 400.0
±0.3°
32.0 to 752.0
E
-240.0 to -200.0
±0.6°
-400.0 to -328.0
-200.0 to -100.0
±0.4°
-328.0 to -148.0
-100.0 to 850.0
±0.3°
-148.0 to 1562.0
850.0 to 1000.0
±0.4°
1562.0 to 1832.0
R
-13.3 to 250.0
±1.4°
250.0 to 750.0
±0.8°
482.0 to 1382.0
750.0 to 1600.0
±0.7°
1382.0 to 2192.0
1600.0 to 1767.8
±0.8°
2192.0 to 3214.0
S
-18.3 to 100.0
±1.4°
100.0 to 400.0
±1.0°
212.0 to 752.0
400.0 to 1700.0
±0.8°
752.0 to 3092.0
1700.0 to 1767.8
±0.9°
3092.0 to 3214.0
B
315.6 to 550.0
±2.0°
600 to 1022.0
550.0 to 900.0
±1.3°
1022.0 to 1652.0
900.0 to 1150.0
±0.9°
1652.0 to 2102.0
1150.0 to 1820.0
±0.8°
2102.0 to 3308.0
(Cold Junction Accuracy not included)
Page 6
Accuracy
T/C
Range
Material
±0.9°
+Iron
-Connstantan
±0.7°
Jacket
±0.5°
±0..7°
±1.4°
+ Chromel®
-Alumel®
±0.7°
Jacket
±0.9°
±2.2°
+Copper
-Constantan
±1.3°
Jacket
-58.0 to 32.0
±0.7°
±0.5°
±1.1°
+Chromel
-Constantan
±0.7°
Jacket
±0.5°
±0.7°
-1.0 to 482.0
±2.5°
+Pt/13Rh
-Platinum
±1.4°
Jacket
±1.3°
±1.4°
-1.0 to 212.0
±2.5°
+Pt/10Rh
-Platinum
±1.8°
Jackrt
±1.4°
±1.6°
±3.6°
+Pt/30Rh
-Pt/6Rh
±2.3°
Jacket
±1.6°
±1.5°
Configure the Calibrator
ISA/ANSI
Color
POWER SWITCH to "SOURCE" or
Move
White
"READ".
Red
Black
MODEL 51#B V#.##
DOUBLE CLICK
FOR CONFIGURATION
Yellow
Red
Setup
Yellow
Double click the
unit is on and the following displays will appear
for 15 seconds:
Blue
> EXIT
Red
TEMP UNITS
Blue
T/C
AUTO OFF
Turn the
DIAL KNOB to move through the
Purple
menu. Press the
Red
between OFF and ON or to scroll through the
Purple
settings.
Black
Red
Simulating thermocouples requires the use of
Green
thermocouple or extension grade thermocouple
wire.
Plug thermocouple wires into the female
miniature thermocouple connector mounted in
Black
the top end of the housing.
Red
Green
The OMEGA CL540A has two banana jacks
mounted in the top end of the housing. These
are not temperature compensated and are to be
used only for millivolt signals.
Grey
Red
Grey
Configuration
EXIT MENU - exits this menu immediately and
saves any changes. Menu will automatically exit
after 15 seconds of inactivity (countdown timer is
displayed).
TEMP UNITS - pressing the knob will toggle
between °C and °F.
T/C -
DIAL KNOB
520: pressing the knob will toggle between the
factory configured T/C Type and mV.
521: pressing the knob will cycle through T/C
DIAL KNOB at any time the
types J, T, E, K, R, S, B, N, G, C, D, L (J-DIN), U
(T-DIN), Platinel II and mV.
AUTO OFF - If AUTO OFF is ON, the unit will
15
turn off after 30 minutes of inactivity to save
°C
battery life. If AUTO OFF is OFF the unit will stay
K
on until the POWER SWITCH is moved to the off
ON
position.
Note: All settings are remembered even with the
DIAL KNOB to toggle
power off. Removing the batteries resets the
values to factory defaults.
Connections
Device
Thermocouple Wire
Requiring
Calibration
CL510AZ
HI
HI
322.1
TYPE K
TYPE K °C
°C
THERMOCOUPLE SOURCE
TYPES
J
T
E
K
R
S
B
N
G
C
D
L
U
P
m V
SOURCE
HI
SET
OFF
LO
Push & Turn
for
Fast Dialing
Push & Hold
to
Store/Ramp
Double Click
for Options
(Optional Boot Shown)
Page 3