Saniclave RS-SC-200
Limitations of Use
Revolutionary Science is not responsible for misuse of the Saniclave 200
1. NEVER attempt to dismantle or modify the Saniclave 200.
2. NEVER leave operating Saniclave 200 unattended.
3. NEVER block drain (located on the inside of the chamber below heating
element).
4. NEVER block steam trap (located on the inside of the chamber, near the top,
directly above drain).
5. NEVER operate the Saniclave 200 with a damaged cord or plug or after the
appliance malfunctions or has been damaged in any manner.
6. NEVER use accessory attachments not recommended by Revolutionary
Science. Their use may cause fire, electrical shock or injury.
7. NEVER place the Saniclave 200 or any part of the Saniclave 200 on or near a
hot gas or electric burner or other heating source.
8. NEVER use the Saniclave 200 or any purpose other than the intended use.
9. ALWAYS close the door properly before operating.
10. NEVER attempt to force the door open or release pressure while sterilizer is
in operation.
ALWAYS use distilled water. The sensor detects ion content in the wa-
11.
ter. If distilled water is too pure, the autoclave may not detect the pres-
ence its presence. Salt may be added to the water to increase detecta-
bility. Class III distilled water is recommended.
12. NEVER reuse water left in the exhaust tank. Re-used water may contain
endotoxins. Drain water tank and refill chamber with clean, distilled water
before each cycle.
13. NEVER attempt to modify the plug to fit an outlet other than the required 3-
slot, grounded outlet. The enclosed 3 prong plug is equipped with a ground-
ed type 3 conductor wire to further ensure the safe operation of the Sani-
clave 200.
14. NEVER immerse the autoclave, cord, or plug in water or operate on wet
surfaces.
15. NEVER fill the exhaust tank with water (See page 3 for location of exhaust
tank and cap). The Saniclave 200 must be filled by pouring water directly
into the chamber.
16. ALWAYS load the Saniclave 200 with instruments compatible with steam
sterilization at 121 degrees Celsius for at least 30 minutes. Always contact
the instrument's manufacturer if you are unsure that this sterilization
technique is suitable for instruments being sterilized.
17. ALWAYS inspect the silicone seal before each use for dirt, grime or cracks.
18. NEVER clean the Saniclave 200 with organic solvents like thinner or benzene.
It will damage the surfaces.
19. ALWAYS test regularly with Geobacillus stearothermophilus. See page 14
for sterilization monitoring and control for more information.
20. This sterilizer has not been validated for double pouching.
R E V O L U T I O N A R Y S C I E N C E
5